58
z
When replacing, use only four same
specified ones.
Take out the battery if you don’t operate the
meter for a long time.
Step 1:
Remove the test leads and Charger
before replacing batteries or fuse, and turn off
the meter.
Step 2:
Remove the protector as shown in
Figure 20.With a standard blade hand
screwdriver, turn each battery door screw a
quarter counterclockwise to remove the
battery door.
z
Step 3:
Replace with four new AAA alkaline
batteries under the instructions shown on the Figure 20 Replacing batteries and fuses
battery door. Or replace the blown fuses with same type F1 (50mA/250V), F2 (50mA/250V) or F3
(63mA/250V).
Step 4:
Reinstall and tighten the battery door, put on the protector before using the meter.
Содержание KM-CAL-907