34
8.2 Measuring DC Voltage
Step 1:
Make sure the lead cables for measurement are not connected to the measuring instrument under
test.
Step 2:
Using the
〔
V
〕
key, select DC Voltage measurement function.
Step 3:
Connect the lead cables for measurement to the measuring terminals of the measuring instrument
under test.
Step 4:
Using the
〔
RANGE
〕
key, select a desired range from 200mV, 5V, 50V. The selected function and the
measured value and unit shall be shown on the LCD.
8.3 Measuring DC Current
Step 1:
Make sure the lead cables for measurement are not connected to the measuring instrument under
test.
Step 2:
Using the
〔
mA
〕
, select DC Current measurement function.
Step 3:
Connect the lead cables for measurement to the measuring terminals of the measuring instrument
under test.
Step 4:
The selected function and the measured value and unit shall be shown in the LCD.