28
Tips:
z
The frequency range could only be changed by pressing
〔
RANGE
〕
key in the frequency set mode.
z
The frequency value and range could be changed when the frequency source function is both in ‘’ON ‘’ or
‘’OFF‘’ state.
7.5 Sourcing Number of Pulses
The calibrator can source a preset number of pulse signal responding to the preset frequency and amplitude.
Step 1:
Using the
〔
PULSE
〕
key, select pulse source function. The LCD shows the default value10 Hz.
Step 2:
Using the
〔
RANGE
〕
key, select a desired frequency range from 100Hz, 1KHz, 10KHz. The selected
function and the default range source value and unit shall be shown on the LCD.
Step 3:
Set the output value digit by digit using each pair of
〔
S
〕
/
〔
T
〕
output setting keys.
Each pair of
〔
S
〕
/
〔
T
〕
keys corresponds to each digit of the LCD reading. Each press of the
〔
S
〕
/
〔
T
〕
key increases or decreases the digit. Increasing the digit from 9 or decreasing it from 0 causes the digit to
overflow or underflow, allowing you to set the output value without interruption. Holding down the
〔
S
〕
/
〔
T
〕
key continuously changes the digit . And the value won’t change if it is increased or decreased to the Maxim or
Minimum value.