32
8.1 Connecting Cables to Terminals
For DC voltage(mV), continuity measurement (Figure 8)
Step 1:
Connect the black lead cable for measurement to
the“COM” input terminal and the red lead cable to
the ”mV” input terminal.
Step 2:
Connect the other end of the cable to the measuring
terminals of equipment under test while making sure the
polarities are correct.
Figure 8 Measuring DC voltage(mV), continuity
For DC voltage(V), Frequency measurement (Figure 9)
Step 1:
Connect the black lead cable for measurement to the
“COM” input terminal and the red lead cable to the ”VHz” input
terminal.
Step 2:
Connect the other end of the cable to the measuring
terminals of equipment under test while making sure the polarities
are correct.
For DC current signal (Figure 10)
Step 1:
Connect the black lead cable for measurement to the
“COM” input terminal and the red lead cable to the ”mA” input Figure 9 Measuring DC voltage(V), frequency
terminal.