49
Measurement Adjustment Operation
Please undertake the adjustment following the sequence and points listed in Table 4.
Table 4 Adjustment Points of Measurement Functions
Range Adjustment
Point
Remarks
0 F
DCV_200mV 200mV
- Notice: replace input
terminal
DCV_5V 5V
-
DCV_50V 50V
-
DCmA_50mA 50mA
-
FREQ_500Hz
500Hz
-
Plus 3V square wave
* Applying reference input signals from the calibration standard as listed in the above table.
Tips:
z
You can also select only the range in need of readjustment to adjust it separately.
Turn on the meter; press the
(MODE)
key while simultaneously holding down the
〔
ON
〕
key into the source
adjusting state, and then press the
〔
MODE
〕
key into the measurement adjusting state, LCD shows “CAL”
symbol on the assistance districts part of the LCD and the reference value and unit on the main districts part of
the LCD.