Ⅳ
Use of rotors beyond the lifetime may lead to breakage of the rotor.
If the rotor is used continuously even after the lifetime of rotors has expired, should
the rotor get damaged, the main unit of the centrifuge suddenly may start to rotate;
this could result in an accident causing injury or death.
WARNING
Usable Rotor
Lifetime of rotors
Lifetime of rotors is 7 years after the delivery.
When 7 years have passed after the delivery, discontinue operation of the centrifuge to replace the
rotor with a new one.
Earlier replacement, however, is required if any corrosion, lowered strength, flaw or deform due to
incorrect operation is detected on the rotor.
In such case, contact your local dealer and be sure to have the rotor checked before reusing it.
WARNING
(1)
Do not use any rotors other than those specified in (2) below.
If rotors other than those specified are used, the rotors may be broken, resulting in
a serious accident.
(2)
The rotors that can mount as of May 2005 are as follows:
This information is subject to addition or change.
For information after June 2005, please contact your local dealer.
Swinging Bucket Rotor
Plate Rotor
RS-240
RS-1004
RMP-23