4-6
Circuit protector
4-6.
Inspection of Circuit Protector
The circuit protector is automatically activated when the electric circuit failure or
overcurrent occur.
In such a case, the white portion of the circuit protector, located at the bottom left on the
rear side of the centrifuge, will stick out.
Recovery procedure is as follows :
(1)
Turn off the power switch.
(2)
After the rotor stops completely, open the lid and check the rotor, following
the procedure given in [3 - 3. Opening and Closing the lid 2. Opening the lid during
power failure] on Page 3 - 4.
(3)
Close the lid , and then push the sticking out white portion until it becomes unseen.
(4)
Turn on the power switch.
(5)
Re - start operation and see if it works.
If the circuit protector is activated again, contact your local dealer.
Section 4 Service