5-6
●
The rotor or metal tubes has been found to be damaged or corroded.
●
A burnt smell comes out of the equipment.
●
You receive a minor electrification when you touch the equipment body
by a naked hand.
●
Some other abnormality or failure has been found occurring.
When operation centrifuge is interrupted by some abnormality, failure or
for repairs, always turn
off
the
“
POWER
”
switch and attach
“
DO NOT
USE
”
stickers to the rotor and the centrifuge before contacting your
nearest dealer.
NOTE
Note 1:
Read
[2-4. Power Requirements] on page 2-2.
3: Read section 6 [Rotor] on page 6-1.
If the error still remains even after the troubleshooting is made and the action has been taken, or
in case the following phenomena are occurring, stop the operation.
Section 5 Troubleshooting
Problem
Checkpoint
Action taken
6. Excessive vibration.
or the error <E1.> lit.
Are buckets and tubes correctly
distributed? (Note 3)
Distribute buckets and tubes
correctly.When the error recurs,
contact your nearest dealer for
inspection.
Do the buckets swing smoothly?
Spray lubricant to buckets and
trunnion pins.
Are the rotor knob or the rotor
locking nut loose?
(Note 3)
Fasten the rotor knob or the rotor
locking nut securely.
Check if the rotor is not loose.
Fasten the rotor locking bolt or
the rotor locking nut securely.
7. An abnormal noise
sounds.
Are buckets and tubes correctly
distributed? (Note 3)
Distribute buckets and tubes
correctly.
Do the buckets swing smoothly?
Spray lubricant to buckets and
trunnion pins.