5-1
The equipment is inoperative while any of the alarm indicators is turned on.
NOTE
“
TIME"display
[Ex.]
Control panel
min
sec
TIME
Cont
’
d. on next page.
Alarm
Indicators
Checkpoint
Action taken
E0
Setting speed alarm
An error occurs immediately after
the "START" key is pressed when
the speed is more than 4,000 rpm
(this may happen when centrifugal
force after changing rotation
radius is changed to speed).
Pressing the
“
START
”
key
makes the centrifuge auto-correct
the speed to 4,000 rpm.
Section 5
Troubleshooting
5-1.
Alarm Indicators
(1)
When trouble occurs during operation, the
“
TIME" display flashes to indicate the alarm.
(2)
When any of the alarms become activated during operation (when the rotor is spin-
ning), the rotor immediately decelerates and stops.