3-8
min
sec
min
sec
FUNCTION
NOTE
●
In case that the operation is stopped before the end of the period set by the
timer and is resumed, the operation period after resumption is the remained
period of the operation.
If the "SPEED/TIME" knob is pressed before resumption of operation, the
operation period returns to the originally set period.
●
The centrifuge can be started at the time when the setting value is flashing at
"TIME" display.
●
When you want to change the setting during the operation, press the
“
FUNCTION" key on the control panel after changing the setting, and the
flashing of the display is stopped.
The operation becomes shifted to the setting that have been changed.
●
The speed and the timer can be set in this sequence or vice versa.
As for the change of the setting, please refer to page 3-15.
3-5.
Setting the Timer
(1)
Pressing the "FUNCTION" key 2 times, the
“
TIME
”
display flashes and indicates the
current setting.
The timer is able to be set up.
(2)
Pressing
“
SPEED/TIME" knob changes the settable range of centrifuging time as des-
cribed below. Set to the range as required.
Factory default
,
the range is set minutes
.
③
HOLD
It will continually operate regardless of
the timer. When set to hold, both "min"
and "sec" lamp will be turn off.
(3)
Turning the
“
SPEED/TIME" knob changes
the indication on the display.
Set to the time required.
(4)
Press
“
START" key on the control panel.
The operation will be started.
When started, the display will diminish and stop at point 0.
Section 3 Operation
②
Setting seconds
①
Setting minutes
Settable range
1-99 minutes
Setting unit
1 minute
Settable range
1-99 seconds
Setting unit
1 second
min
sec
min
sec
min
sec