3-4
The lid is locked in following
cases and will not open even if
the
〈
OPEN
〉
button is pressed.
・
When the rotor is spinning.
・
When there is a power failure
or the power is turned off.
Do not open the lid while the rotor is spinning.
When a power failure occurs during operation of the centrifuge, the rotor
will naturally decelerate to stop its rotation. If the lid is unlocked forcibly
before the rotor stops, the person can be entangled in the rotor, resulting
in a serious injury or death.
WARNING
START
ST
O P
TIM E
mi n
sec
S
PEED
100r pm
100g
ME
MORY
12 3
ACC
DEC
F
UNCTION
“
OPEN
”
button
Open
NOTE
3-3.
Opening and Closing the lid
[1]
Turning on the power and opening the lid
(1)
Turn on the the
“
POWER" switch to turn the power on.
The
“
POWER" switch locates on the front side panel of
the centrifuge.
(2)
The
〈
STOP
〉
lamp on the control panel will be lit.
(3)
Pressing the
〈
OPEN
〉
button the lid can be opened.
[2]
Opening the lid during power failure
(1)
Turn off the
“
POWER
”
switch.
(2)
By using a flat-blade screwdriver, remove the lid
extreme release cap on the lower part of the left
side of the unit.
While pulling the string straight, push
“
OPEN
”
button. (Do not pull the string too strong.)
Then the lid opens.
Section 3 Operation
String
ST
AR
T
ST
OP
TIME
SP
EE
D
FU
NC
T
I O
N