Ⅰ
Do not perform dry-heat or auto-
clave sterilization with tempera-
tures higher than those specified.
Otherwise, the rotor may deterio-
rate and cause an accident.
(2)
Maximum speed
(4)
Hazardous material
General Notes
Since large electrical and mechanical energies are present on the centrifuge and rotor,
reasonable care is required for their handling.
Otherwise, failure may occur resulting in property damages or fatal physical injuries.
In order to prevent them from happening, be sure to follow the instruction given below.
Be sure to follow them.
(1)
Maximum load
WARNING
(5)
While the centrifuge is in operation
(6)
L
id
Do not exceed the maximum load
of rotor and bucket.
The rotor or bucket used beyond
the allowable load level can be
damaged, thereby causing an ac-
cident.
Keep the rotor and bucket speed
below the maximum speed.
Excessive speed may cause dam-
age to the rotor, bucket and the
centrifuge. The maximum speed
depends on the rotor and bucket
strength.
Do not modify, nor use unspeci-
fied parts.
Unauthorized retrofit of the centri-
fuge, rotor or bucket, or use of un-
authorized parts for them can re-
sult in accidents.
Whilethe centrifugeis in operation,
do not stand closer than 30 cm to
avoid asecondary accident.
Do not open the lid when the rotor
is spinning.
Physical contact with the spinning
rotor or bucket may cause serious
injury.
(9)
Damaged, corroded, rusted or deformed
(8)
Rotor and drive shaft during rotation
Do not touch rotor and drive shaft
during the rotation.
Physical contact with the spinning
rotor or drive shaft may cause
serious injury.
Discontinue use of the equipment
when its rotor or the buckets
found to have been damaged,
corroded, rusted or deformed.
Otherwise failure may occur.
(7)
Sterilization
Do not put any hazardous materi-
al (explosive, chemically active,
organic, or radiation containing
material, or material contaminated
by pathogenic microorganisms) as
a sample of the centrifuge and do
not place it closer than 30 cm to
avoid a secondary accident
should the centrifuge accidentally
rotate and contact the material.
(3)
Modification and unspecified parts