5-3
“
TIME"display
[Ex.]
Control panel
min
sec
TIME
The equipment is inoperative while any of the error indicators is turned on.
NOTE
Cont
’
d. on next page.
Error
Indicators
Checkpoint
Action taken
E1
Imbalance error
This indicator flashes and the
rotor stops rotation when
excessive vibration is
detected.
①
Balance the load (samples or bucket).
②
Remove dust and dirt from the sliding
surface and apply grease so that the bucket
on the swing rotor swings up smoothly.
Contact your local dealer for inspection when
this error continues to flash even after taking
the above measures.
E2
Lid open error
This indicator lights when
the lid is open.
Close the lid completely.
When the error recurs, contact your nearest
dealer for inspection.
E3
Inverter Error
This indicator flashes when
abnormal signal comes out
from the inverter.
After the rotor stops spinning, turn off power
supply once and turn on the power after about
30 seconds.
When the error recurs, contact your nearest
dealer for inspection.
5-2.
Error Indicators
(1)
When trouble occurs during operation, the
“
TIME" display flashes to show the error. The
error indicators are mentioned in the table below.
(2)
When any of the error indicators turn on during operation (when the rotor is spinning),
the rotor immediately decelerates and stops.
Section 5 Troubleshooting