3-13
100rpm
×
100 g
×
min
sec
SPEED
TIME
100rpm
×
100 g
×
min
sec
SPEED
TIME
FUNCTION
[2]
Setting the "SLOW" deceleration
The characteristics have become as shown in the graph.
You can set the natural deceleration starting speed
N
by the following method.
Factory default
,
it is set as 0 rpm
.
The setting range of "N" is from 0 to 4,000 rpm.
(1)
Pressing the "SPEED/TIME" knob while holding down the "FUNCTION" key.
The used number of times of the centrifuge is displayed.
(2)
Pressing the "SPEED/TIME" knob several times and let indicate "
"on the "TIME"
display.
The "SPEED" display indicates the current setting of the natural deceleration speed.
Unit : rpm
Refer to Page 3-11.
[
Ex.
]
Setting is the 0 rpm.
(3)
Turn the "SPEED/TIME" knob to set the natural deceleration speed.
The setting range of "N" is from 0 to 4,000 rpm.
[
Ex.
]
Setting the 300 rpm.
(4)
Press the "SPEED/TIME" knob while holding down the "FUNCTION" key.
The display returns to the previous Speed and Time indication after the setting ( Natural
deceleration speed) is saved.
Section 3 Operation
N
1000
N
1000
(A)N =
1,000 rpm or more
(B)N =
1,000 rpm or less
r
pm
r
pm
speed
It changes to the SLOW
characteristic from RAPID
at 1000rpm.
RAPID
Natural
deceleration
SLOW
TIME
TIME
speed
RAPID
Natural
deceleration