53
The displayed measurement function can be changed in the measurement mode by pressing
↑
key.
The setting of the display output is demonstrated for the example of displaying the mass flow
rate in kg/h.
Starting from the display of measurement values, the following programming steps are
necessary.
Begin from measuring mode.
After pressing
→
key the display shows :
0000.0000 (kg)/min
This display means that mass flow will be displayed in units of kg/min to a resolution of 4
decimal places.
The brackets around 'kg' indicate the cursor position, these characters will be flashing on the
display. The flashing value can now be changed with the
↑
key. By pressing the
→
key the
cursor moves to 'min' which starts to flash.
This dimension may now also be changed by using the
↑
key. After pressing the
→
key again,
the cursor moves to the output format of the numeric value which can then be edited.
To change the display to kg/h with 5 decimal places, follow the following procedure.
Key
Display
line 1
line2
→
Fct. (1).0
OPERATOR
2x
↑
Fct. (3).0.
INSTALL
→
Fct. 3.(1).0.
BASE DATA
↑
Fct. 3.(2).0.
DISPLAY
→↑↑
Fct. 3.2.(3).
MASS FLOW
Key
Display
line 1
line 2
0000.0000
(kg)/min
→
0000.0000
kg/(min)
↑
0000.0000
kg/(h)
→
0000(.)0000
kg/h
↑
00000(.)000
kg/h
↑
000000(.)00
kg/h
↑
0000000(.)0
kg/h
↑
00000000(.)
kg/h
↑
0(.)0000000
kg/h
↑
00(.)000000
kg/h
↑
000(.)00000
kg/h
↵
Fct. 3.2.(3).
MASS FLOW