76
6.14.4 Primary head type and tube parameters (CF1-9)
The primary head type and tube parameters are factory set and should not normally be
changed by the customer. They are only normally needed if the converter has to be replaced in
the field. In this case, the new converter must be programmed with the correct primary head
type and transducer-specific parameters.
In the case of remote mounted converters, particularly, the customer may wish to verify that
the correct converter is connected to a given transducer. This may be accomplished by
comparing the primary type and CF5 (GK) programmed in the converter with the
corresponding data printed on the transducer data tag.
To view these parameters, begin from the measuring mode:
To view the other transducer-specific parameters, proceed to Fct. 3.9.0 TUBE PARAMS.
Check the values programmed against the Data Plate and/or Calibration Certificate data. If
the data is incorrect, it can be adjusted in Fct. 3.9.0 by first unprotecting the parameters via
Code 4 password (see Fct. 3.8.8).
6.14.5 Location
It is possible to program each unit with an individual identification number. This is especially
useful if the 'SMART' option is being used. To set the location number :
Begin from measuring mode:
Use the
↑
key to change the character under the cursor. The sequence is :
A to Z, 0 to 9, +, -, *, /, =, SPACE.
Use the
→
key to move the cursor to the next character.
Press
↵
key when completed.
Key
Display
line 1
line 2
→↑↑
Fct. (3).0
INSTALL
→→
4x
↑
Fct. 3.1.(5)
USER DATA
→
check for correct primary type
1.5E - 1500P
↵↑
Fct. 3.1.(6)
CF5
→
Value
CF5
Compare value with data tag
4x
↵
Key
Display
line 1
line 2
→↑↑
Fct. (3).0
INSTALL
→
7x
↑
Fct. 3.(8).0
USER DATA
→↑↑↑
Fct. 3.8.(4)
LOCATION
→
(M)FC 081
Factory Setting