Fct. No.
Text
Description and settings
3.4.0
PULS.OUTP. P
Submenu 3.4.0 Frequency output P
3.4.1
FUNCTION P
Function frequency output P
* OFF (Output = 0V DC)
* MASS FLOW (Frequency output 0 to MAX Freq. Hz =
Mass Flow in range: MIN. FLOW to MAX FLOW as set
in Fct. 3.4.3 and 3.4.4)
* MASS TOTAL(1 pulse = fixed mass as set in Fct 3.4.2)
* DENSITY (Frequency output 0 to MAX Freq. Hz =
Density in range: MIN.DENSITY to MAX.DENSITY as
set in Fct. 3.4.3 and 3.4.4)
* TEMPERAT. (Frequency output 0 to MAX Freq. Hz =
Temperature in range: MIN. TEMP to MAX. TEMP as
set in Fct. 3.4.3 and 3.4.4)
* VOLUME.FLOW(Frequency output 0 to MAX Freq. Hz
= Volume flow in range: MIN. V.FLOW to MAX.
V.FLOW as set in Fct. 3.4.3 and Fct. 3.4.4)
* VOL.TOTAL(1 pulse = fixed volume as set in Fct 3.4.2)
Solute flow
Solute total
Concentration parameters if
Conc. by mass
option installed. See separate
Conc. by volume
instruction manual.
* DIRECTION (Negative flow gives output of 0 volts DC,
Positive flow gives output of +V volts DC)
3.4.2
PULSE/MASS
Mass per pulse value for function TOTAL MASS
or
PULSE/VOL.
Volume per pulse value for function VOL. TOTAL
or
PULSE/TIME
Maximum frequency value for functions MASS FLOW,
DENSITY, TEMPERATUR and VOLUME.FLOW or
CONC.OPTIONS. Not accessible for functions OFF and
DIRECTION.
3.4.3
MIN.VALUE
Value of measured quantity that corresponds to
or
MIN. FLOW,
0 Hz output
or
MIN. DENSITY,
or
MIN. TEMP.
or
MIN. V.FLOW
or
CONC.OPTIONS
or
PULSE.WIDTH
For functions MASS TOTAL, VOL.TOTAL OR
SOL.TOTAL. Not accessible for functions OFF and
DIRECTIONS
3.4.4
Full Scale
Value of measured quantity that corresponds to
or
MAX. FLOW,
Max. Frequency
or
MAX. DENSITY
or
MAX TEMP.
or
MAX V.FLOW
or
CONC.OPTIONS
Not accessible for functions OFF, DIRECTION, TOTAL
MASS TOTAL, or VOL. TOTAL
}