44
Fct. No.
Text
Description and settings
3.9.0
TUBE PARAMS
Submenu 3.9.0 Transducer calibration and
compensation parameters*
3.9.1
Fgw
CF1
Drive frequency - water: from Calibration Certificate
3.9.2
Fcw
CF2
Coriolis frequency - water: from Calibration
Certificate
3.9.3
Fgl
CF3
Drive frequency - air: from Calibration Certificate
3.9.4
Fcl
CF4
Coriolis frequency - air: from Calibration Certificate
3.9.5
GK
CF5
Transducer mass flow calibration constant: from
Calibration Certificate
3.9.6
LIN
CF6
Linearity adjustment: from Calibration Certificate
3.9.7
Tcl
CF7
Mass flow temperature compensation: from
Calibration Certificate
3.9.8
Tc0
CF8
Mass flow temperature compensation at zero flow:
from
Calibration Certificate
3.9.9
TcD
CF9
Density temperature compensation: from Calibration
Certificate
3.9.10
D.REF.WATER
Use
↑
key to select between two modes, then press
↵
to
select
submenu and then
↑
for choice of parameters
* MEAS.VALUE
CALIB. NO
CALIB. YES
* SET VALUE
freq Hz
temp
°
C (
°
F)
density g/cm
3
3.9.11
D.REF.AIR
Use
↑
key to select between two modes, then press
↵
to
select
submenu and then
↑
for choice of parameters
* MEAS.VALUE
CALIB. NO
CALIB. YES
* SET VALUE
freq Hz
temp
°
C (
°
F)
density g/cm
3
* Many, but not necessarily all, compensation parameters are shown on instrument data
plate.
These menus, except 3.9.10 and 3.9.11, are protected by the Code 4 password; see
Fct. 3.8.8.