Chapter 17 Troubleshooting
17-21
E
30D2
Temperature
error
When the rapid cooling outlet 30 second
moving average temperature is compared
to 50
°
C, an error occurs when the
temperature is 5
°
C higher when Ready
and 10
°
C higher when in PRINT.
1.
Check HPRO OFFSET and mechanical control offset
2.
Check the connection between the H-DRVboard and
mechanical control board
3.
Check connection between mechanical control
board and HPRO unit
4.
Check conductance of the power cable
5.
Check voltage
6.
Replace H-DRV board
7.
Replace heat process unit
8.
Replace mechanical control board
9.
Replace the wire bundle between the H-DRVboard
and mechanical control board
10.
Replace wire bundle between mechanical control
board and HPRO unit
C
30E2
Caution
1 Minute average value temperature
upper limit error
(supply)
1.
Check the connection to the internal temperature
sensor mechanical control board (CN22)
2.
Replace mechanical control board
3.
Replace the internal temperature sensor
C
30E3
Caution
1 Minute average value temperature
upper limit error
(supply)
F
30ED
HPRO Error
Check the connector disconnect signal for
the internal temperature sensor every 0.1
seconds, and an error occurs if an
anomaly is detected
1.
Check the connection of connector CN22 on the
mechanical control board
2.
Replace mechanical control board
3.
Replace the internal temperature sensor
E
3A0E
HPRO Error
Deodorization FAN rotation error
1.
Check connection between mechanical control
board and deodorization FAN
2.
Replace the deodorization fan
3.
Replace mechanical control board
4.
Replace wire bundle between deodorization FAN
and mechanical control board
E
3B0E
HPRO Error
Supply cooling FAN rotation error
1.
Check connection between mechanical control
board and supply cooling FAN
2.
Replace the supply cooling FAN
3.
Replace mechanical control board
4.
Replace wire bundle between supply cooling FAN
and mechanical control board
E
3C0E
HPRO Error
Heat insulation unit FAN rotation error
1.
Check heat insulation unit FAN connector
2.
Replace heat insulation unit FAN
3.
Replace mechanical control board
4.
Replace wire bundle between heat insulation FAN
and mechanical control board
C
3DB2
Caution
Standard temperature upper limit error
(heating inlet)
Replace HPRO unit
C
3DC2
Caution
Standard temperature upper limit error
(slow cool outlet)
C
3DD2
Caution
Standard temperature upper limit error
(rapid cooling outlet)
C
3DB3
Caution
Standard temperature lower limit error
(heating inlet)
C
3DC3
Caution
Standard temperature lower limit error
(slow cool outlet)
C
3DD3
Caution
Standard temperature lower limit error
(rapid cooling outlet)
E
4110
Pickup not
required
Could not pickup at CH1 pickup
operation (film does not reach the
position regulation FILM detection
sensor).
•
Shake the tray, and re-insert the film
•
Replace pickup unit
•
Check the connection and operation of the pickup
motor and mechanical control board
•
Check the connection and operation of the pickup
evacuation motor and mechanical control board
•
Check the connection and operation of the FILM
detection sensor and mechanical control board
•
Replace mechanical control board
E
4120
The operation
was not
performed
normally
The pickup evacuation sensor did not
switch ON during CH1 pickup operation.
•
Replace pickup unit
•
Check the connection and operation of the pickup
evacuation motor and mechanical control board
•
Check the connection and operation of the pickup
evacuation sensor and mechanical control board
•
Replace mechanical control board
Error
Level
Error
Code
Error
Message (Display
Characters)
Error contents
Error Support: Service
Содержание Drypro 832
Страница 2: ......
Страница 12: ......
Страница 15: ...Chapter 1 Safety Precautions and Warnings 1 3 5 Laser Power Label No Precautions Warning Labels ...
Страница 22: ......
Страница 28: ......
Страница 74: ......
Страница 118: ......
Страница 228: ......
Страница 242: ......
Страница 260: ...Chapter 14 Setting and Adjustment 14 4 6 Click the PRINT button The ST13A PATTERN is printed Example ST13A PATTERN ...
Страница 263: ...Chapter 14 Setting and Adjustment 14 7 12 Click the PRINT button The ST13B PATTERN is printed Example ST13B PATTERN ...
Страница 265: ...Chapter 14 Setting and Adjustment 14 9 16 Click the PRINT button The ST13C PATTERN is printed Example ST13C PATTERN ...
Страница 388: ......
Страница 498: ......
Страница 521: ...18 23 18 5 Block Diagram ...
Страница 527: ...18 29 18 7 Printing Operation Load Timing Chart Standard 1 ch ...
Страница 529: ......