Chapter 11 Replacing Loading Parts in Main Unit
11-25
11.9 Replacing Control Cooling Fan
Replace the fan unit that cools the control unit.
Work outline
The number of personnel and the estimated work hours required to replace loading parts is as follows:
Requirements
The procedure for replacing loading parts requires the following tools
1
Remove the LAN cover and film ejection tray with
reference to
“
Removal Procedure (Page 3-
10)”
in
“3.3.2 Removing/Installing Film Ejection
Tray”
.
2
Remove the connector cover.
3
Remove connectors from MEC IF (CCN5),
exposure IF (CCN6), and power supply (JP52).
Caution
Be sure to turn off power and remove the power cable from the electrical outlet before removing
or installing the exterior cover or replacing loading parts.
Number of Personnel
Work Hours
1
10 minutes
No.
Tools
No.
Tools
1
Various types of screwdrivers
2
Tray (container for screws and parts removed)
3
Fan unit (for replacement)
Caution
When removing the connector, do not pull by
the wires, but hold the tab on the connector
with your fingers.
Pulling by the wires too tightly can cause a
break.
Содержание Drypro 832
Страница 2: ......
Страница 12: ......
Страница 15: ...Chapter 1 Safety Precautions and Warnings 1 3 5 Laser Power Label No Precautions Warning Labels ...
Страница 22: ......
Страница 28: ......
Страница 74: ......
Страница 118: ......
Страница 228: ......
Страница 242: ......
Страница 260: ...Chapter 14 Setting and Adjustment 14 4 6 Click the PRINT button The ST13A PATTERN is printed Example ST13A PATTERN ...
Страница 263: ...Chapter 14 Setting and Adjustment 14 7 12 Click the PRINT button The ST13B PATTERN is printed Example ST13B PATTERN ...
Страница 265: ...Chapter 14 Setting and Adjustment 14 9 16 Click the PRINT button The ST13C PATTERN is printed Example ST13C PATTERN ...
Страница 388: ......
Страница 498: ......
Страница 521: ...18 23 18 5 Block Diagram ...
Страница 527: ...18 29 18 7 Printing Operation Load Timing Chart Standard 1 ch ...
Страница 529: ......