Chapter 17 Troubleshooting
17-19
E
3092
Temperature
error
When the slow cool C1 30 second
moving average temperature is compared
to 100
°
C, an error occurs when the
temperature is 5
°
C higher when Ready
and 10
°
C higher when in PRINT.
1.
Check HPRO OFFSET and mechanical control offset
2.
Check the connection between the H-DRVboard and
mechanical control board
3.
Check connection between mechanical control
board and HPRO unit
4.
Check conductance of the power cable
5.
Check voltage
6.
Replace H-DRV board
7.
Replace heat process unit
8.
Replace mechanical control board
9.
Replace the wire bundle between the H-DRVboard
and mechanical control board
10.
Replace wire bundle between mechanical control
board and HPRO unit
E
3093
Temperature
error
When the slow cool C1 30 second
moving average temperature is compared
to 100
°
C, an error occurs when the
temperature is 5
°
C lower when Ready
and 10
°
C lower when in PRINT.
1.
Check HPRO OFFSET and mechanical control offset
2.
Check whether the short-circuit breaker (domestic)
or circuit protector (domestic) is OFF
3.
Check the connection between the H-DRVboard and
mechanical control board
4.
Check connection between mechanical control
board and HPRO unit
5.
Check conductance of the power cable
6.
Check voltage
7.
Replace H-DRV board
8.
Replace heat process unit
9.
Replace mechanical control board
10.
Replace the wire bundle between the H-DRVboard
and mechanical control board
11.
Replace wire bundle between mechanical control
board and HPRO unit
E
3094
Temperature
error
Irrespective of the status of the machine,
check the temperature of slow cool C1
every 1 second, and if the temperature is
continually 147
°
C for 5 seconds, an error
occurs
1.
Check HPRO OFFSET and mechanical control offset
2.
Check the connection between the H-DRVboard and
mechanical control board
3.
Check connection between mechanical control
board and HPRO unit
4.
Check conductance of the power cable
5.
Check voltage
6.
Replace H-DRV board
7.
Replace heat process unit
8.
Replace mechanical control board
9.
Replace the wire bundle between the H-DRVboard
and mechanical control board
10.
Replace wire bundle between mechanical control
board and HPRO unit
E
30A2
Temperature
error
When the slow cool C2 30 second
moving average temperature is compared
to 100
°
C, an error occurs when the
temperature is 5
°
C higher when Ready
and 10
°
C higher when in PRINT.
E
30A3
Temperature
error
When the slow cool C2 30 second
moving average temperature is compared
to 100
°
C, an error occurs when the
temperature is 5
°
C lower when Ready
and 10
°
C lower when in PRINT.
1.
Check HPRO OFFSET and mechanical control offset
2.
Check whether the short-circuit breaker (domestic)
or circuit protector (domestic) is OFF
3.
Check the connection between the H-DRVboard and
mechanical control board
4.
Check connection between mechanical control
board and HPRO unit
5.
Check conductance of the power cable
6.
Check voltage
7.
Replace H-DRV board
8.
Replace heat process unit
9.
Replace mechanical control board
10.
Replace the wire bundle between the H-DRVboard
and mechanical control board
11.
Replace wire bundle between mechanical control
board and HPRO unit
E
30A4
Temperature
error
Irrespective of the status of the machine,
check the temperature of slow cool C2
every 1 second, and if the temperature is
continually 147
°
C for 5 seconds, an error
occurs
1.
Check HPRO OFFSET and mechanical control offset
2.
Check the connection between the H-DRVboard and
mechanical control board
3.
Check connection between mechanical control
board and HPRO unit
4.
Check conductance of the power cable
5.
Check voltage
6.
Replace H-DRV board
7.
Replace heat process unit
8.
Replace mechanical control board
9.
Replace the wire bundle between the H-DRVboard
and mechanical control board
10.
Replace wire bundle between mechanical control
board and HPRO unit
Error
Level
Error
Code
Error
Message (Display
Characters)
Error contents
Error Support: Service
Содержание Drypro 832
Страница 2: ......
Страница 12: ......
Страница 15: ...Chapter 1 Safety Precautions and Warnings 1 3 5 Laser Power Label No Precautions Warning Labels ...
Страница 22: ......
Страница 28: ......
Страница 74: ......
Страница 118: ......
Страница 228: ......
Страница 242: ......
Страница 260: ...Chapter 14 Setting and Adjustment 14 4 6 Click the PRINT button The ST13A PATTERN is printed Example ST13A PATTERN ...
Страница 263: ...Chapter 14 Setting and Adjustment 14 7 12 Click the PRINT button The ST13B PATTERN is printed Example ST13B PATTERN ...
Страница 265: ...Chapter 14 Setting and Adjustment 14 9 16 Click the PRINT button The ST13C PATTERN is printed Example ST13C PATTERN ...
Страница 388: ......
Страница 498: ......
Страница 521: ...18 23 18 5 Block Diagram ...
Страница 527: ...18 29 18 7 Printing Operation Load Timing Chart Standard 1 ch ...
Страница 529: ......