13
Installation and initial startup
BlueShadow Detector 40D Instructions V7647
4.4 Inserting the flow cell
The supply configuration of this detector only includes a test cell (dum-
my cell) from KNAUER, which does not have connections for solvent . It is
used, for example, to check the lamp intensity, as a dirty flow cell could
distort this value . Before using the device with solvents, the test cell must,
there fore, be removed, and a flow cell must be installed .
If necessary, the optical path length has been set .
Procedure
Process
Figure
1.
Unscrew the knurled-head
screws
3
of the cover plate
4
. Hold the flow cell securely
with your hand during this
procedure .
2.
Pull out the slide with the flow
cell
5
towards the front .
3.
Remove the flow cell upward .
4.
Position the new flow cell on
top of the slide . Keep hol-
ding the flow cell . Make sure
that the alignment pin of the
detector is aligned with the
corresponding hole in the
flow cell (see figure 4) .
5.
Push the slide back into the
detector .
6.
Screw the knurled-head
screws back into the cover
plate
4
and tighten .
7.
Connect the incoming and
outgoing capillaries at the
inlet
1
and out let
2
of the
measuring cell .
Fig. 3:
Inserting the flow cell
Connect the capillaries .
Prerequisites
1
2
3
4
5
3
Next step