VIQ-Installation-Qualification
Version 2 .1 Page
2/3
Installation Qualification (IQ)
Packing list, list of hardware and software
Initial startup
Position
Device/accessory
Order number
Comment
Step
Action
Result
Passed Failed
Comment
1
Identify the device .
The device is correct .
2
Check the device for
transport damage .
No transport damage
is observed .
3
Check the scope of
delivery .
The scope of delivery
is complete .
4
Connect the device to
the power supply .
The device is connected .
5
Connect all loose
parts (e .g . capillaries,
tubings, measuring
head) .
The device is fully
assembled and ready
to use .
6
Switch on the device .
The device starts (operating
noise) . The power LED or
the display lights up .
Please compare the article numbers listed on the packing list to the provided products . Note deviations here: