56
Maintenance and care
BlueShadow Detector 40D Instructions V7647
Note:
If the cleaning of the light guide does not have the desired effect,
the light guide has to be replaced .
Mount the flow cell onto the detector .
8.5 Replacing the fiber optics
UV light will cause the fiber optics to become blind with time (solariza-
tion), making them no longer suitable for use .
Note:
Observe the following regarding the use of fiber optics:
Do not touch the ends of the fiber optics with your fingers, as this
could falsify the measurement .
Handle the fiber optics with care and avoid impacts .
Move the fiber optics carefully without using pressure or bending it .
Detector is switched off .
Cap fittings for fiber optics
Procedure
Process
Figure
1.
To detach the fiber optics
from the flow cell, unscrew
the fittings of the fiber optics
manually
1
.
Fig. 19
Fiber optic fittings on
flow cell
2.
Seal the fiber optic connec-
tors with caps
2
.
Fig. 20
Cap for optic connector
3.
Manually unscrew the fiber
optics
3
from the detector .
4.
To mount new fiber optics
onto the flow cell, repeat step
1-3 in reverse order .
Fig. 21
Fiber optics on the detector
The fiber optics have been replaced .
Next step
Prerequisite
Tools
Result