49
Troubleshooting
BlueShadow Detector 40D Instructions V7647
System message
”Invalid time in time
table”
Correct the time entry .
”Invalid time table
index“
Change the entry in the program line .
L
”Link is loaded“
First unload the link then change the link or
delete it .
“Link is running”
Wait until the link has been completed, then
change the link or delete it .
N
”No link available“
Create a link and edit it .
”No link available,
pls edit link first“
Create a link and edit it .
”No time table to
start“
Edit the data by using the keypad .
”Not enough space
to store link“
Check the detector . Check the number of pro-
gram lines . A maximum of 100 program lines
are possible .
”Not enough space
to store program“
Check the detector . Check the number of pro-
gram lines . A maximum of 100 program lines
are possible .
P
”Program does not
exist“
Create a program .
”Program is
running“
Quit program or wait until program has been
completed .
T
”This link is used in
Wake up“
First quit or delete wakeup program (wu = Wake
Up), then edit or delete link .
”This program is
used in a link“
First pause or delete the link, then edit or delete
data by using the keypad .
”This program is
used in Wake up“
First quit or delete wakeup program (wu = Wa-
ke Up), then edit or delete data by using the
keypad .
”Time already
exists“
Correct the time entry .
”Time table is not
active“
The device is in Standalone mode, no program
is running . If you try to quit a non-existent pro-
gram sequence, this message appears .
”Time table is not
loaded“
First load the program, then start the program .
”Time table line is
empty“
Edit the program line .