23
Installation and initial startup
BlueShadow Detector 40D Instructions V7647
4.7.2 Connecting the pin header
To control one device through another, the pin header is used .
To use remote control, you have to connect cables to the pin header .
The single ports are used to exchange control signals .
The device is switched off .
The power plug is disconnected from the device .
Depressor tool
Electronic defect
Connecting cables to the multi-pin connector of a switched on device
cau ses a short circuit .
Turn off the device before connecting cables .
Pull the power plug .
Electronic defect
Electrostatic discharge can destroy the electronics .
Wear a protective bracelet against electrostatic discharge and ground .
2
3
1
Fig. 11:
Depressor tool
Process
1.
Insert the depressor tool
3
in an upper small opening at the front
of the pin header
1
.
2.
Insert the cable into the opening
2
underneath the inserted de-
pressor tool .
3.
Pull out the depressor tool .
Check whether the cables are tightly attached . Press the pin header onto
the connector . Finish the installation . Bring the device into operation .
Prerequisites
Tools
Next steps
Содержание BlueShadow Detector 40D
Страница 1: ...Detector 40D Instructions Document no V7647 HPLC HPLC...
Страница 71: ...Appendix Installation Qualification IQ...