11
BlueShadow Detector 40D Instructions V7647
4. Installation and initial startup
Before you determine the operation site, read the chapter “Technical
data” (see chapter 11 on page 59) . There you will find all device-speci-
fic information on power supply, ambient conditions and humidity .
Note:
The intended use be ensured only if the requirements for ambient
conditions of the operating environment are met .
4.1 Unpacking
Process
1.
Place the packaging in such a way that the lettering on the label is
in the correct position .
2.
Check the packaging, the device and the accessories for transport
damage .
3.
Check the scope of delivery . In the event of incomplete delivery,
contact technical support immediately .
4.
When lifting, carrying or moving the device, grab the unit only
from below on the sides . Do not hold onto front cover or leak tray,
as these parts are loosely attached to the device .
Keep the included packing list for repeat orders .
Keep the original packaging for safe storage or transportation .
4.2 Ambient conditions
4.2.1 Operation site
Observe the following requirements for the operation site so that the
measurement results are not influenced:
Place on a firm, level and straight surface .
Protect against direct sunlight .
Do not expose to air drafts such as air conditioning systems .
Do not set up the to other machines that cause floor vibrations .
Keep from high frequency sources .
Ensure adequate ventilation (see „4 .2 .3 Space requirements“, p . 12) .
Avoid temperature fluctuations (see „4 .2 .2 Ambient temperature“, p .
4.2.2 Ambient temperature
If the ambient temperature of the device is abruptly changed (e .g . when
it is installed in a cooling chamber), condensation will form inside the
device and may cause damage to the device . Allow the device to accli-
mate for 3 h, before it is connected to the power supply and taken into
operation .
Procedure
Next steps
Содержание BlueShadow Detector 40D
Страница 1: ...Detector 40D Instructions Document no V7647 HPLC HPLC...
Страница 71: ...Appendix Installation Qualification IQ...