ELS-304
8-2 Programming Mode
8.2
Meas. Commands
Function
Command
Description
Memo
Load curr. meas.
MC{NR1} range 0〜2
*1
0:AUTO range
readback value:real number(##.###)
1:curr.High range
2:curr.Low range
MV{NR1} range: 0〜2
*1
0:AUTO range
readback value:real number(##.###)
1:volt.High range
2:volt.Low range
Power meas.
MW
Load terminal volt.×load curr.
calsulated power result is back
readback value:real number(##.###)
Define meas. display
DS{NR1}
range: 0〜1
*2
0:define upper Meas.display
1:define lower Meas.display
Mes. Fixed mode
MF{NR1}
range 0〜1
meas. command selected last will be
0:free run meas.
executed repeatedly.
1:single meas.
HZ{NR1}
range 0〜1
sampling rate of the AD for
0:50Hz
measurement can be setup.
1:60Hz
Notes
{ } : you can NOT mot in this par NR1・・・・・・ integral value
*1: Setting range and the meas. range of load section can not be setup independently
but should be set as identical..
*2: When Power switch is turned on, lower display is defined when GPIB measured (default Setting)
Display where upper or lower are
effective is define when GPIB
measured
Volt. at the load
teminal meas.
meas. frequency setting
Содержание ELS-304
Страница 1: ...ELS 304 High SPEED Electronic Load OPERATION MANUAL Engineering Company SERIES ...
Страница 12: ...ELS 304 ...
Страница 22: ...ELS 304 2 8 Connections ...
Страница 50: ...ELS 304 6 4 Operations for Meas Modes ...
Страница 72: ...ELS 304 8 18 Programming Mode ...
Страница 76: ...ELS 304 9 4 Specifications ...
Страница 80: ......