ELS-304
Operations of Load Modes
5-11
5.10
Short (SHORT) Mode
In this mode, load terminals are shortened
Load current can flow up to the rated 120A or setup current limit whichever smaller, but the
current range is fixed to 120A .
(1) Press
the
“rotary knob” to set mode select condition. and then turn the knob until “ST” is
displayed as shown below.
(2)
Input setting value by “ten keys”, and then press “ENT” key to confirm the setting value.
(3) Press
“ON/OFF” key to select ON to start flowing the load current up to the rated 120A
or setup current limit whichever smaller.
Fig. 5-10-1
Operation in SHORT mode
7
4
1
0
.
2
5
8
9
6
3
CE
ON/OFF
ENT
Mode Select
Set Value
confirm
Load ON
The change of a set value can be passed