ELS-304
Alarm
7-1
7.
Alarms
The protection functions and alarms of this product are described.
When an Alarm occurs, beep sound rings, an error message is displayed, and the Load will
promptly be set OFF to the no-load condition.
7.1
Over Current Limit
When this alarm works, the Load Current is limited to protect the Load section of this
product. The Lord Current will be limited by the 110% of the Current Limit Vale. To
setup the Current Limit, refer to “
5.4
Curr. limit setting
”.
If the Current limit is not setup, the Current Limit value will be 120A as the default
value.
7.2
Over Voltage Alarm
When the load volt exceeds the tolerance value, this alarm operates.
In High range, when Vin
≧
31.2V(Typ), in Low range when Vin
≧
4.16V(Typ) this alarm
is issued.
Fig.7-2-1
Display of Over Voltage Alarm
・
This is for the alarm only, there is no protection for the Load Section of this
product.
・
When this alarm works, if there is questionable odor or abnormal noise, please
stop using this product immediately and call us or our distributor for the repair
WARNING
Содержание ELS-304
Страница 1: ...ELS 304 High SPEED Electronic Load OPERATION MANUAL Engineering Company SERIES ...
Страница 12: ...ELS 304 ...
Страница 22: ...ELS 304 2 8 Connections ...
Страница 50: ...ELS 304 6 4 Operations for Meas Modes ...
Страница 72: ...ELS 304 8 18 Programming Mode ...
Страница 76: ...ELS 304 9 4 Specifications ...
Страница 80: ......