ELS-304
2-2
Connections
2.2
Load Terminal
(1) Remote Sense Cable
Never fail to connect a Remote Sense Cable that should be shielded or twisted. If the
Remote Sense cable is NOT used, Volt Meas. at the Load Terminal and Data in each
Load Modes (CR, CV, CP) using the Volt, Reference may have errors.
・
Before you connect this cable, turn off the power of the Target Device.
・
This product have the protection resistor to protect against the reverse
connection at the Load Terminals.
Because of this, the Volt. Meas. value may be lower due to the Volt. Drop of the
resistor. There is no such resistor at he Remote Sense Terminals.
TARGET DEVICE
+
-
SG
(1)
(2)
(3)
WARNING
NOTE