Keysight PD1000A Power Device Measurement System Startup Guide
41
IV and CV Test Requirements
IV and CV Device Characterization
Create the IV and CV and CV-IV Switching Setups in the Library
This section uses the UHC Selector Input Selection.xkd file that you downloaded
to a memory stick. See “Download Setup and Configuration Files” on page 19.
1
For the B1506A, insert the memory stick into the front panel.
2
In the tabs at the left, select
Application Test.
3
Then select
Library > Import....
Show
All Files.
4
Locate and import the
UHC Selector Input Selection.xtd
file.
5
Click the
UHC Selector Input Selection
icon.
Setup the IV to CV Switch
1
Specify the HVSMU slot. (See Figure 15 above and refer to the Modules tab in
on page 37. Typically, the HVSMU slot is Slot 7.)
2
For the SMU Slot Number, enter
0
(zero).
3
Click the
Save
tab (right side of screen) to save it to My Favorite.
4
Right click the icon and rename it as
IV to CV
. IMPORTANT: use this exact file
name as shown, including the spaces between words and capitalization.
Figure 15
Configure the IV to CV and CV to IV Switch Parameters
Содержание PD1000A
Страница 1: ...Keysight PD1000A Power Device Measurement System for Advanced Modeling Startup Guide ...
Страница 2: ......
Страница 4: ......
Страница 8: ...viii Keysight PD1000 Power Device Measurement System Startup Guide ...
Страница 96: ...94 Keysight PD1000A Power Device Measurement System Startup Guide PD1000A System Specifications ...
Страница 104: ...102 Keysight PD1000A Power Device Measurement System Startup Guide Glossary IGBT Terms ...
Страница 107: ......