66
Keysight PD1000A Power Device Measurement System Startup Guide
On State and Off State S-Parameter Characterization
Test Requirements
General Test Setup
1
Connect two PD1000A Bias T networks to your PD1000A Test Fixture as
shown below:
2
Refer to Figure 33 below. From the B2902A Precision SMU, connect the dual
banana plug cables to the two PD1000A Bias T networks as follows:
a
Connect four (two red and two black) dual banana plug cables to the Bias
T that provides bias voltage to the gate of the Device Under Test (DUT).
b
The other ends of the dual banana plug cables connect to the B2902A
front panel Force High, Force Low, Sense High, and Sense Low connectors.
c
Connect four (two red and two black) dual banana plug cables to the Bias
T that provides bias to the drain/collector of the DUT.
d
The other ends of the dual banana plug cables connect to the B2902A rear
panel Force High, Force Low, Sense High, and Sense Low connectors.
Figure 32
Connecting Two Bias Ts to the Test Fixture (TO-247 fixture shown,
cable not shown)
The Bias Ts and the Test Fixture must be supported on a table or
bench top.