28
Keysight PD1000A Power Device Measurement System Startup Guide
IV and CV Device Characterization
B150xA Specification Conditions
The measurement and output accuracy are specified under the conditions listed
below. Note: The SMU measurement and output accuracy are specified at the
output terminals in the test fixture except for capacitance measurement that is
specified at the output terminals of the MFCMU.
–
Temperature: 23 ±5 °C.
–
Humidity: 20 to 70%, No condensation.
–
Self-calibration after a 40 minute warm-up is required.
–
Room temperature change less than ±1 °C after self-calibration execution.
(Note: This does not apply to the MFCMU).
–
Measurement made within one hour after self-calibration execution. (Note:
This does not apply to the MFCMU).
–
Calibration period: 1 year.
–
SMU integration time setting: 10 PLC (1 nA to 1 A range, voltage range),
200 μs (20 A range) Averaging of high-speed ADC: 128 samples per 1 PLC.
–
SMU filter: ON for MPSMU.
–
The accuracy of the drain output current measurement specification is not
guaranteed until 20 seconds after a voltage change.
B150xA Operating Conditions
The B1506A or B1505A PDA must to be used under the conditions listed below.
–
Temperature: +5 to +40 °C
–
Humidity: 20 to 70%, No condensation
–
When used with InTEST Thermal plate:
-- Temperature: +5 to +30 °C
-- Humidity: 20% to 70%, No condensation
-- The B1505A requires the Keysight N1265A text fixture to work with the
InTEST platform.
Important:
Always use the latest firmware in the analyzer. Firmware
www.keysight.com/find/B1505A
Use EasyEXPERT Software version 6.1.1742.9630 (or later).
Software updates are available at:
www.keysight.com/find/EasyEXPERT.
Содержание PD1000A
Страница 1: ...Keysight PD1000A Power Device Measurement System for Advanced Modeling Startup Guide ...
Страница 2: ......
Страница 4: ......
Страница 8: ...viii Keysight PD1000 Power Device Measurement System Startup Guide ...
Страница 96: ...94 Keysight PD1000A Power Device Measurement System Startup Guide PD1000A System Specifications ...
Страница 104: ...102 Keysight PD1000A Power Device Measurement System Startup Guide Glossary IGBT Terms ...
Страница 107: ......