Keysight PD1000A Power Device Measurement System Startup Guide
97
Configure the E5810B Gateway for the InTEST Controller
Configure the InTEST Thermal Platform Controller
9
Click the
IDN?
button to read the identification string for the InTEST
Thermal Platform Controller.
Configure the Remote GPIB Interface in Connection Expert
In the Keysight IO Libraries Suite Connection Expert utility, make certain that
Connection Expert can communicate with the InTEST Thermal Platform
controller. Identify the VISA address of the Controller. Use this VISA address
when configuring the PD1000A Power Device Measurement System for
Advanced Modeling Power Control Software.
1
Launch Keysight Connection Expert from the Start menu.
2
If a Remote GPIB Interface is not listed in the Connection Expert My
Instruments
list
,
select
+Add > Remote GPIB Interface
to display the Remote
Interface dialog.
3
Enter the IP address from the E5810B into the Hostname or IP Address
field.
Figure 51 IDN?
String for the InTEST Thermal Platform Controller