Keysight PD1000A Power Device Measurement System Startup Guide
IV and CV Device Characterization
IV and CV Test Requirements
Setup the CV to IV Switch.
1
Specify the HVSMU slot. (See Figure 15 above and refer to the Modules tab in
“Configure EasyEXPERT” on page 37. Typically, the HVSMU slot is Slot 7.)
2
Enter
1 f
or the SMU Slot Number.
3
Click the
Save
tab (right side of screen).
4
Right-click and rename as
CV to IV
. IMPORTANT: use this exact file name as
shown, including the spaces between words and capitalization.
Testing the Switch Setup
1
Select the
IV to CV
Favorite.
2
Click the
Recall
button.
3
Click the green
Play
arrow or
RUN > Single
.
4
Click the
Close
button. You should hear clicking in the test chamber.
5
Select
Tools > System > Configuration
.
6
Select the
UHC Expander / Fixture
tab. The Input SMU field should be empty.
7
Select the
CV to IV
Favorite.
8
Click the
Recall
button.
9
Click the green
Play
arrow or
RUN > Single
.
10
Click the
Close
button. You should hear clicking in the test chamber.
11
Click
Tools > System > Configuration
.
12
Click the
UHC Expander / Fixture
tab. The Input SMU field should have the SMU
name.
Содержание PD1000A
Страница 1: ...Keysight PD1000A Power Device Measurement System for Advanced Modeling Startup Guide ...
Страница 2: ......
Страница 4: ......
Страница 8: ...viii Keysight PD1000 Power Device Measurement System Startup Guide ...
Страница 96: ...94 Keysight PD1000A Power Device Measurement System Startup Guide PD1000A System Specifications ...
Страница 104: ...102 Keysight PD1000A Power Device Measurement System Startup Guide Glossary IGBT Terms ...
Страница 107: ......