1002
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
30
:WAVeform Commands
Peak detect display mode is used to detect glitches for time base settings of
500 us/div and slower. In this mode, the oscilloscope can sample more data than
it can store and display. So, when peak detect is turned on, the oscilloscope scans
through the extra data, picks up the minimum and maximum for each time bucket,
then stores the data in an array. Each time bucket contains two data sample.
The array is transmitted over the interface bus linearly, starting with time bucket 0
proceeding through time bucket n-1, where n is the number returned by the
:WAVeform:POINts? query (see
). In each time bucket, two values are
transmitted, first the minimum, followed by the maximum. The first pair of values
corresponds to the time bucket at the leftmost side of the screen. The last pair of
values corresponds to the time bucket at the far right side of the screen. In
:ACQuire:TYPE PEAK mode (see
), the value returned by the
:WAVeform:XINCrement query (see
) should be doubled to find the time
difference between the min-max pairs.
HRESolution Data
The high resolution (
smoothing
) mode is used to reduce noise at slower sweep
speeds where the digitizer samples faster than needed to fill memory for the
displayed time range.
Data Conversion
Word or byte data sent from the oscilloscope must be scaled for useful
interpretation. The values used to interpret the data are the X and Y references, X
and Y origins, and X and Y increments. These values are read from the waveform
preamble. Each channel has its own waveform preamble.
In converting a data value to a voltage value, the following formula is used:
voltage = [(data value - yreference) * yincrement] + yorigin
If the :WAVeform:FORMat data format is ASCii (see
are converted internally and sent as floating point values separated by commas.
In converting a data value to time, the time value of a data point can be
determined by the position of the data point. For example, the fourth data point
sent with :WAVeform:XORigin = 16 ns, :WAVeform:XREFerence = 0, and
:WAVeform:XINCrement = 2 ns, can be calculated using the following formula:
time = [(data point number - xreference) * xincrement] + xorigin
This would result in the following calculation for time bucket 3:
time = [(3 - 0) * 2 ns] + 16 ns = 22 ns
In :ACQuire:TYPE PEAK mode (see
), because data is acquired in
max-min pairs, modify the previous time formula to the following:
time=[(data pair number - xreference) * xincrement * 2] + xorigin
Data Format for Transfer
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...