:SEARch Commands
26
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
841
:SEARch:SERial:LIN:MODE
Command Syntax
:SEARch:SERial:LIN:MODE <value>
<value> ::= {ID | DATA | ERRor | FRAMe | FSIGnal}
The :SEARch:SERial:LIN:MODE command selects the type of LIN information to
find in the Lister display:
• ID — searches for a frame ID.
• DATA — searches for a frame ID and data.
• ERRor — searches for errors.
• FRAMe — searches for symbolic frames.
• FSIGnal — searched for symbolic frames and a signal values.
Frame IDs are specified using the :SEARch:SERial:LIN:ID command.
Data values are specified using the :SEARch:SERial:LIN:PATTern:DATA command.
Frames, signals, and signal values are specified using the
:SEARch:SERial:LIN:SYMBolic:FRAMe, :SEARch:SERial:LIN:SYMBolic:SIGNal, and
:SEARch:SERial:LIN:SYMBolic:VALue commands. LIN symbolic data files are
loaded (recalled) using the :RECall:LDF[:STARt] command.
Query Syntax
:SEARch:SERial:LIN:MODE?
The :SEARch:SERial:LIN:MODE? query returns the currently selected mode.
Return Format
<value><NL>
<value> ::= {ID | DATA | ERR | FRAM | FSIG}
See Also
•
, “:SEARch Commands,” starting on page 785
•
•
":SEARch:SERial:LIN:PATTern:DATA"
•
•
":SEARch:SERial:LIN:SYMBolic:FRAMe"
•
":SEARch:SERial:LIN:SYMBolic:SIGNal"
•
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...