614
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
25
:SBUS<n> Commands
:SBUS<n>:CAN:TRIGger
Command Syntax
:SBUS<n>:CAN:TRIGger <condition>
<condition> ::= {SOF | EOF | IDData | DATA | FDData | IDRemote
| IDEither | ERRor | ACKerror | FORMerror | STUFferror | CRCerror
| SPECerror | ALLerrors | BRSBit | CRCDbit | EBActive | EBPassive
| OVERload | MESSage | MSIGnal | FDMSignal}
The :SBUS<n>:CAN:TRIGger command sets the CAN trigger on condition:
Condition
Front-panel name
Description
Filter by
ID*
SOF
SOF - Start of Frame
Triggers at the start bit for both data and
overload frames.
EOF
EOF - End of Frame
Triggers at the end of any frame.
X
IDEither
Frame ID
Triggers on any standard CAN (data or
remote) or CAN FD frame at the end of the
11- or 29-bit ID field.
IDData
Data Frame ID (non-FD)
Triggers on standard CAN data frames at
the end of the 11- or 29-bit ID field.
DATA
Data Frame ID and Data
(non-FD)
Triggers on any standard CAN data frame at
the end of the last data byte defined in the
trigger. The DLC of the packet must must
match the number of bytes specified.
FDData
Data Frame ID and Data
(FD)
triggers on CAN FD frames at the end of the
last data byte defined in the trigger. You can
trigger on up to 8 bytes of data anywhere
within the CAN FD data, which can be up to
64 bytes long.
IDRemote
Remote Frame ID
Triggers on standard CAN remote frames at
the end of the 11- or 29-bit ID field.
ERRor
Error Frame
Triggers after 6 consecutive 0s while in a
data frame, at the EOF.
X
ACKerror
Acknowledge Error
Triggers on the acknowledge bit if the
polarity is incorrect.
X
FORMerror Form Error
Triggers on reserved bit errors.
X
STUFferror Stuff Error
Triggers on 6 consecutive 1s or 6
consecutive 0s, while in a non-error or non
overload frame.
X
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...