:ACQuire Commands
:ACQuire:RSIGnal
Command Syntax
:ACQuire:RSIGnal <ref_signal_mode>
<ref_signal_mode> ::= {OFF | OUT | IN}
The :ACQuire:RSIGnal command selects the reference signal mode. A common
reference signal can be used by multiple instruments to synchronize their
timebases.
• The OFF mode disables the oscilloscope's REF I/O MMCX connector. The
oscilloscope's internal 10 MHz reference signal is used, and there is no
timebase synchronization.
• The OUT mode outputs the oscilloscope's internal 10 MHz reference signal to
the REF I/O MMCX connector.
• The IN mode synchronizes the oscilloscope's timebase to an external 10 MHz
square or sine wave signal that is input to the MMCX connector labeled REF
I/O. The amplitude must be from - 5 dBm to 17 dBm (356 mVpp to 4.48 Vpp).
Query Syntax
:ACQuire:RSIGnal?
The :ACQuire:RSIGnal? query returns the current reference signal mode.
Return Format
<ref_signal_mode><NL>
<ref_signal_mode> ::= {OFF | OUT | IN}
See Also
•
CAUTION
Do not apply more than 20 dBm Max (6.32 Vpp Max) at the REF I/O MMCX connector
on the front panel or damage to the instrument may occur.
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...