:SBUS<n>:SENT:PPULse
Command Syntax
:SBUS<n>:SENT:PPULse {{0 | OFF} | {1 | ON} | SPC}
The :SBUS<n>:SENT:PPULse command specifies whether there is a pause pulse
between Fast Channel Messages:
• OFF — There is no pause pulse between Fast Channel Messages.
Note that a SENT serial bus with no pause pulse is never idle. This means,
during normal operation, the fast channel decode line will show a continuous
stream of packets, with a new packet opening as soon as the previous one has
closed.
• ON — Pause pulses are added between Fast Channel Messages so that frames
come at a regular interval.
If there is a pause pulse (and
Pause Pulse
is on), idle time is shown between
messages.
• SPC (Short PWM Code) — In SENT SPC, there are no pause pulses. Instead, the
message event is triggered by the master when it wants to receive data. SENT
SPC ends the transmission after the CRC so it almost appears as if there is a
pause pulse from the end until the next master trigger.
Query Syntax
:SBUS<n>:SENT:PPULse?
The :SBUS<n>:SENT:PPULse? query returns the pause mode setting.
Return Format
<setting><NL>
<setting> ::= {0 | 1 | SPC}
See Also
•
•
•
•
•
•
•
":SBUS<n>:SENT:SIGNal<s>:DISPlay"
•
":SBUS<n>:SENT:SIGNal<s>:LENGth"
•
":SBUS<n>:SENT:SIGNal<s>:MULTiplier"
•
":SBUS<n>:SENT:SIGNal<s>:OFFSet"
•
":SBUS<n>:SENT:SIGNal<s>:ORDer"
•
":SBUS<n>:SENT:SIGNal<s>:STARt"
•
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...