:CALibrate Commands
9
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
249
:CALibrate:OUTPut
Command Syntax
:CALibrate:OUTPut <signal>
<signal> ::= {TRIGgers | MASK | WAVEgen | WGEN1 | NFC | TSOurce | OFF}
Note: WAVE and WGEN1 are equivalent.
The CALibrate:OUTPut command sets the signal that is available on the Aux Out
MMCX connector:
• TRIGgers — pulse when a trigger event occurs. The output level is 0-5 V into an
open circuit, and 0-2.5 V into 50
Ω
.
• MASK — signal from mask test indicating a failure.
• WAVEgen, WGEN1 — waveform generator sync output signal. This signal
depends on the :WGEN<w>:FUNCtion setting:
• NFC — This option is available in the Near Field Communication (NFC) trigger
mode when the ATRigger (Arm & Trigger) trigger event is selected (see
on page 949). The ATRigger trigger event lets you arm
the oscilloscope on one event and then trigger on a second event or after a
specified timeout period if the second event does not occur.
When :CALibrate:OUTPut is NFC and the specified event arms, the Aux Out
goes high. The oscilloscope waits until a second event is found or until the
specified timeout period expires and then triggers.
•
For NFC-A, the second event is SDD_REQ.
•
For NFC-B, the second event is ATTRIB.
•
For NFC-F, the second event is ATR_REQ.
When the oscilloscope triggers, the Aux Out line goes low.
Waveform
Type
Sync Signal Characteristics
SINusoid,
SQUare,
RAMP, PULSe,
SINC,
EXPRise,
EXPFall,
GAUSsian
The Sync signal is a TTL positive pulse that occurs when the waveform rises above
zero volts (or the DC offset value).
DC, NOISe,
CARDiac
N/A
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...