308
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
13
:DISPlay Commands
:DISPlay:GRATicule:TYPE
Command Syntax
:DISPlay:GRATicule:TYPE <type>
<type> ::= {FULL | MVOLt | IRE}
The :DISPlay:GRATicule:TYPE command sets the graticule (grid) type.
When the TV trigger type is selected (see
vertical scaling of at least one displayed channel is 140 mV/div, the
:DISPlay:GRATicule:TYPE command lets you select from these grid types:
• FULL — the normal oscilloscope grid.
• MVOLt — shows vertical grids, labeled on the left, from -0.3 V to 0.8 V.
• IRE — (Institute of Radio Engineers) shows vertical grids in IRE units, labeled on
the left, from -40 to 100 IRE. The 0.35 V and 0.7 V levels from the MVOLt grid
are also shown and labeled at the right. When the IRE grid is selected, cursor
values on the display are also shown in IRE units. However, cursor values via
the remote interface are not in IRE units.
The MVOLt and IRE grid values are accurate when the vertical scaling is
140 mV/division and the vertical offset is 245 mV.
Query Syntax
:DISPlay:GRATicule:TYPE?
The :DISPlay:GRATicule:TYPE? query returns the graticule (grid) type setting.
Return Format
<type><NL>
<type> ::= {FULL | MVOL | IRE}
See Also
•
•
•
•
•
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...