526
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
22
:MTESt Commands
:MTESt:SCALe:X1
Command Syntax
:MTESt:SCALe:X1 <x1_value>
<x1_value> ::= X1 value in NR3 format
The :MTESt:SCALe:X1 command defines where X=0 in the base coordinate system
used for mask testing. The other X-coordinate is defined by the
:MTESt:SCALe:XDELta command. Once the X1 and XDELta coordinates are set, all
X values of vertices in the mask regions are defined with respect to this value,
according to the equation:
X = (X *
Δ
X) + X1
Thus, if you set X1 to 100 ms, and XDELta to 100 ms, an X value of 0.100 is a
vertex at 110 ms.
The oscilloscope uses this equation to normalize vertices. This simplifies
reprogramming to handle different data rates. For example, if you halve the period
of the waveform of interest, you need only to adjust the XDELta value to set up the
mask for the new waveform.
The X1 value is a time value specifying the location of the X1 coordinate, which will
then be treated as X=0 for mask regions coordinates.
Query Syntax
:MTESt:SCALe:X1?
The :MTESt:SCALe:X1? query returns the current X1 coordinate setting.
Return Format
<x1_value><NL>
<x1_value> ::= X1 value in NR3 format
See Also
•
"Introduction to :MTESt Commands"
•
•
•
•
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...