440
Keysight P9241/42/43A Oscilloscopes SCPI Programmer's Guide
21
:MEASure Commands
:MEASure:DEFine
Command Syntax
:MEASure:DEFine <meas_spec>[,<source>]
<meas_spec> ::= {DELay | THResholds}, for remaining syntax, see:
•
":MEASure:DEFine DELay Command Syntax"
•
":MEASure:DEFine THResholds Command Syntax"
The :MEASure:DEFine command sets up the definition for measurements by
specifying the delta time or threshold values. Changing these values may affect
the results of other measure commands. The table below identifies which
measurement results that can be affected by redefining the DELay specification or
the THResholds values. For example, changing the THResholds definition from the
default 10%, 50%, and 90% values may change the returned measurement result.
:MEASure:DEFine
DELay Command
Syntax
MEASure Command
DELay
THResholds
DUTYcycle
x
DELay
x
x
FALLtime
x
FREQuency
x
NWIDth
x
OVERshoot
x
PERiod
x
PHASe
x
PREShoot
x
PWIDth
x
RISetime
x
VAVerage
x
VRMS
x
NOTE
The DELay portion of the MEASure:DEFine command has been deprecated, and you should
instead use the :MEASure:DELay:DEFine command to specify delay measurement parameters.
The delay measurement now allows an edge occurrence of zero (0) to specify automatic edge
selection. The limitation of this command is that you cannot specify a negative slope with an
occurrence count of zero (0).
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...