The WAVeform subsystem is used to transfer data to a controller from the
oscilloscope waveform memories. The queries in this subsystem will only operate
when the channel selected by :WAVeform:SOURce is on.
Waveform Data and Preamble
The waveform record is actually contained in two portions: the preamble and
waveform data. The waveform record must be read from the oscilloscope by the
controller using two separate commands, :WAVeform:DATA (see
) and
:WAVeform:PREamble (see
). The waveform data is the actual data
acquired for each point in the specified source. The preamble contains the
information for interpreting the waveform data, which includes the number of
points acquired, the format of acquired data, and the type of acquired data. The
preamble also contains the X and Y increments, origins, and references for the
acquired data, so that word and byte data can be translated to time and voltage
values.
Data Acquisition Types
There are four types of waveform acquisitions that can be selected for analog
channels with the :ACQuire:TYPE command (see
): NORMal, AVERage,
PEAK, and HRESolution. When the data is acquired using the :DIGitize command
) or :RUN command (see
), the data is placed in the
channel buffer of the specified source.
Once you have acquired data with the :DIGitize command, the instrument is
stopped. If the instrument is restarted (via the programming interface or the front
panel), or if any instrument setting is changed, the data acquired with the :DIGitize
command may be overwritten.You should first acquire the data with the :DIGitize
command, then immediately read the data with the :WAVeform:DATA? query (see
) before changing any instrument setup.
A waveform record consists of either all of the acquired points or a subset of the
acquired points. The number of points acquired may be queried using
:ACQuire:POINts[:ANALog]? (see
Helpful Hints:
n/a
:WAVeform:YORigin?
<return_value> ::= y-origin in
the current preamble in NR3
format
n/a
:WAVeform:YREFerence?
<return_value> ::= y-reference
value in the current preamble in
NR1 format
Table 138
:WAVeform Commands Summary (continued)
Command
Query
Options and Query Returns
Содержание P9241
Страница 1: ...Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 6: ...6 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide...
Страница 36: ...36 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 1 What s New...
Страница 41: ...Setting Up 2 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 41...
Страница 44: ...44 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 2 Setting Up...
Страница 222: ...222 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 7 Root Commands...
Страница 244: ...244 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands...
Страница 256: ...256 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands...
Страница 278: ...278 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 10 CHANnel n Commands...
Страница 328: ...328 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands...
Страница 386: ...386 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands...
Страница 390: ...390 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands...
Страница 416: ...416 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands...
Страница 498: ...498 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 21 MEASure Commands...
Страница 532: ...532 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 22 MTESt Commands...
Страница 572: ...572 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 24 SAVE Commands...
Страница 596: ...596 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 25 SBUS n Commands SBUS n A429 SOURce on page 588...
Страница 900: ...900 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 28 TIMebase Commands...
Страница 1080: ...1080 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 31 WGEN w Commands...
Страница 1142: ...1142 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 34 Error Messages...
Страница 1204: ...1204 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide 37 More About Oscilloscope Commands...
Страница 1344: ...1344 Keysight P9241 42 43A Oscilloscopes SCPI Programmer s Guide Index...