From the remote interface:
To select the trigger output function:
DIG:PIN1:FUNC TOUT
To select the pin polarity:
DIG:PIN1:POL POS
DIG:PIN1:POL NEG
Output relay
Digital Control pins 1 and 2 can be programmed to function as an output relay. All pins are referenced to the Signal
Common pin.
When the output of the power supply is turned off, it is implemented by setting the output to 0 volts and 0.02 amps.
This gives a zero-output voltage without actually disconnecting the output. To disconnect the output, an external
relay must be connected between the output and the load. A TTL signal of either low true or high true is provided to
control an external relay. This signal can only be controlled with the remote command OUTPut:RELay OFF | ON. The
TTL output is available on the Digital IO pin 1 for channel 1 and pin 2 for channel 2. When the OUTPut:RELay state is
“ON” on channel 1, the TTL output of pin 1 is high if the Polarity is positive and pin 1 is low if the Polarity is negative.
The levels are reversed when the OUTPut:RELay state is “OFF”.
From the front panel:
1. Press
Utilities
>
I/O Config
>
Digital IO
to access the Digital I/O window.
2. Select the pin you wish to configure. For example, press
Pin 1
to configure Pin 1.
3. Press
Function
and select
Relay
. Select and program the remaining pins in the same manner. Press
Back
to
return to previous page.
4. Configure the polarity for each of the pin. Press
Polarity Pos
to select Positive and
Polarity Neg
to select
Negative. Select and program the remaining pins in the same manner.
From the remote interface:
To select the trigger output function:
DIG:PIN1:FUNC REL
To select the pin polarity:
DIG:PIN1:POL POS
DIG:PIN1:POL NEG
86
Keysight E36200 Series User's Guide
Содержание E36200 Series
Страница 1: ...User s Guide E36200 Series Autoranging DC Power Supplies ...
Страница 24: ...Dimension diagram E36231A E36232A and E36234A dimensions 24 Keysight E36200 Series User s Guide ...
Страница 34: ...Front panel Rear panel 34 Keysight E36200 Series User s Guide ...
Страница 125: ...Guide d utilisation Série E36200 Alimentations CC à commutation automatique de plage ...
Страница 148: ...Schéma de dimensions Dimensions de E36231A E36232A et E36234A 24 Guide d utilisation de la série Keysight E36200 ...