Perform Calibration
opens the calibration menu. Make sure you have run the self-test before performing any
calibration. Refer to "Calibration Adjustment Procedures" in the Service Guide for details.
Press
Change Passcode
to change the passcode using the numeric keypad. The passcode can be set up to 9 digits.
Auto save On
will automatically save the calibration data upon logout. When
Auto save
is Off, press
Cal Save
to
save the calibration data.
Self-Test
Self Test
verifies proper instrument operation.
Self-Test Procedures
A power-on self-test occurs automatically when you turn on the power supply. This limited test assures you that the
power supply is operational.
Press
Utilities
>
Test / Setup
>
Self Test
to perform the complete self-test of the power supply. It takes
approximately 2 seconds for the self-test to complete.
You can also perform a complete self-test from the remote interface, see the
E36200 Series Programming Guide
for details.
– If the self-test is successful, "
Self test passed
" is displayed on the front panel.
– If the self-test fails, "
!Err
" is displayed on the front panel. Record the error code and message and contact
Keysight support if necessary.
– If the self-test is successful, this indicates a high chance that the power supply is operational.
User Settings
User Settings
specifies user preferences that control how the user interacts with the instrument. These settings are
stored in non-volatile memory.
Sound Settings
Beeper
enables or disables the beeper sound when an error is generated from the front panel or remote interface.
Key Click
enables or disables the click heard when a front-panel key or softkey is pressed.
116
Keysight E36200 Series User's Guide
Содержание E36200 Series
Страница 1: ...User s Guide E36200 Series Autoranging DC Power Supplies ...
Страница 24: ...Dimension diagram E36231A E36232A and E36234A dimensions 24 Keysight E36200 Series User s Guide ...
Страница 34: ...Front panel Rear panel 34 Keysight E36200 Series User s Guide ...
Страница 125: ...Guide d utilisation Série E36200 Alimentations CC à commutation automatique de plage ...
Страница 148: ...Schéma de dimensions Dimensions de E36231A E36232A et E36234A 24 Guide d utilisation de la série Keysight E36200 ...