A power-on self-test occurs automatically when you turn the unit on. This assures you that the instrument is
operational. If self-test fails, or if other operating problems occur with your instrument, the front panel error
indicator (
!Err
) appears at the upper top of the display.
View the error log
Press
Utilities
>
Error
to display the error log. If there are more than 10 errors on the display, press
Next
to scroll to
the next page.
Press
Back
or
[Meter View]
to return to the meter-view display.
– Errors are stored in the order they are received. The error at the end of the list is the most recent error.
– If there are more than 20 errors in the queue, the last error stored is replaced with -350,"Queue overflow". No
more errors are stored until you remove errors from the queue. If there are no errors, the instrument responds with
+0,"No error".
– Except for self-test errors, errors are cleared when exiting the Error Log menu or when cycling power.
If you suspect that there is a problem with the power supply, refer to the "Troubleshooting" section in the
Service
Guide
.
64
Keysight E36200 Series User's Guide