From the remote interface:
To enable current or voltage data logging on outputs 1 and 2:
SENS:DLOG:FUNC:CURR 1,(@1,2)
SENS:DLOG:FUNC:VOLT 1,(@1,2)
You cannot data log output power from the remote interface. To obtain power data, you need to data log both
voltage and current and then calculate the power from the resulting voltage and current data.
To log the minimum and maximum values to the data log file for all enabled outputs:
SENS:DLOG:FUNC:MINM 1
To specify a data log of 1000 seconds on all enabled outputs:
SENS:DLOG:TIME 1000
To specify a sample period of 400 milliseconds between data samples on all enabled outputs:
SENS:DLOG:PER 0.4
To send an immediate trigger signal to the data logger:
TRIG:DLOG
To select the immediate trigger source (triggers the Data Logger immediately when initiated):
TRIG:DLOG:SOUR IMM
To select the rear panel trigger input (all connector pins that have been configured as trigger sources):
TRIG:DLOG:SOUR EXT
To select a BUS trigger source:
TRIG:DLOG:SOUR BUS
Save the data log
Press
File Selection
to choose between internal logging or external logging. Select
Int
to save the data into the
internal memory or
Ext
to save the data into an external USB drive.
102
Keysight E36200 Series User's Guide
Содержание E36200 Series
Страница 1: ...User s Guide E36200 Series Autoranging DC Power Supplies ...
Страница 24: ...Dimension diagram E36231A E36232A and E36234A dimensions 24 Keysight E36200 Series User s Guide ...
Страница 34: ...Front panel Rear panel 34 Keysight E36200 Series User s Guide ...
Страница 125: ...Guide d utilisation Série E36200 Alimentations CC à commutation automatique de plage ...
Страница 148: ...Schéma de dimensions Dimensions de E36231A E36232A et E36234A 24 Guide d utilisation de la série Keysight E36200 ...